Leica Thunder: Difference between revisions

From GI
Jump to navigation Jump to search
Started page
 
No edit summary
 
Line 2: Line 2:
== Manuals and user guides ==
== Manuals and user guides ==


== Biosafety and access ==
The SoRa is located in an ML-1 room. See the [[General Instrumentation Department#Biosafety|Biosafety]] section on rules regarding working in this room, and how to get access.


== Current status ==
== Current status ==
The current status of the microscope can be found [[Current microscope status#Light microscopes|here]] and on the [https://bookings.science.ru.nl/category/general-instrumentation-gi/gi-light-microscopy-image-analysis booking website].
The current status of the microscope can be found [[Current microscope status#Light microscopes|here]] and on the [https://bookings.science.ru.nl/category/general-instrumentation-gi/gi-light-microscopy-image-analysis booking website].
== Technical specifications ==
=== Microscope ===
Automated DMi8 (inverted) with hardware autofocus (AFC) and motorized polarizer.
=== Excitation Lines ===
Lumencor SPECTRA-X 7-line LED with the following available wavelengths:
* 395 nm
* 440 nm
* 470 nm
* 510 nm
* 550 nm
* 640 nm
=== Emission Selection ===
* DFT5 quad-cube for high speed fluorescence imaging (Blue, Green, Red, Far-red, or 100%) enabling specific capture of:
** 420-450 nm (eg. DAPI, Hoechst)
** 506-532 nm (eg. GFP, FITC)
** 581-607 nm (eg. RFP, TRITC, Cy3)
** 666-724 nm (eg. AlexaFluor 633, DRAQ5, Cy5)
* CFP filter cube to capture:
** 460-500 nm (e.g. CFP, TFP, Cerulean)
* YFP filter cube to capture:
** 520-550 nm (e.g. YFP, mVenus)
=== Detection ===
Leica DFC9000 GTC
* 4.2 MP sCMOS camera, 6.5 μm pixel size (19 mm sensor diagonal)
* 82% quantum efficiency, dark current of 0.14e-/p/sec
* ca. 90 frames per second acquisition speed
* extended dynamic range of 33000:1
=== Objectives ===
* 5x/0.12 Dry (working distance 14 mm)
* 10x/0.40 Dry (2.74 mm)
* 20x/0.80 Air (0.40 mm)
* 40x/1.10 Water (0.65 mm)
* 63x/1.20 Water (0.30 mm)
* 63x/1.30 Glycerol (0.28 mm)
=== Stage ===
High speed/precision quantum stage
=== Environmental Control ===
Incubator with temperature controller
=== Software Modules ===
* LAS X 3D for 3D rendering of acquired volumes
* “Navigator” for large tile scan
* “Thunder” for instant or volume computational clearing and improvement of resolution

Latest revision as of 11:43, 29 September 2025

Manuals and user guides

Biosafety and access

The SoRa is located in an ML-1 room. See the Biosafety section on rules regarding working in this room, and how to get access.

Current status

The current status of the microscope can be found here and on the booking website.

Technical specifications

Microscope

Automated DMi8 (inverted) with hardware autofocus (AFC) and motorized polarizer.

Excitation Lines

Lumencor SPECTRA-X 7-line LED with the following available wavelengths:

  • 395 nm
  • 440 nm
  • 470 nm
  • 510 nm
  • 550 nm
  • 640 nm

Emission Selection

  • DFT5 quad-cube for high speed fluorescence imaging (Blue, Green, Red, Far-red, or 100%) enabling specific capture of:
    • 420-450 nm (eg. DAPI, Hoechst)
    • 506-532 nm (eg. GFP, FITC)
    • 581-607 nm (eg. RFP, TRITC, Cy3)
    • 666-724 nm (eg. AlexaFluor 633, DRAQ5, Cy5)
  • CFP filter cube to capture:
    • 460-500 nm (e.g. CFP, TFP, Cerulean)
  • YFP filter cube to capture:
    • 520-550 nm (e.g. YFP, mVenus)

Detection

Leica DFC9000 GTC

  • 4.2 MP sCMOS camera, 6.5 μm pixel size (19 mm sensor diagonal)
  • 82% quantum efficiency, dark current of 0.14e-/p/sec
  • ca. 90 frames per second acquisition speed
  • extended dynamic range of 33000:1

Objectives

  • 5x/0.12 Dry (working distance 14 mm)
  • 10x/0.40 Dry (2.74 mm)
  • 20x/0.80 Air (0.40 mm)
  • 40x/1.10 Water (0.65 mm)
  • 63x/1.20 Water (0.30 mm)
  • 63x/1.30 Glycerol (0.28 mm)

Stage

High speed/precision quantum stage

Environmental Control

Incubator with temperature controller

Software Modules

  • LAS X 3D for 3D rendering of acquired volumes
  • “Navigator” for large tile scan
  • “Thunder” for instant or volume computational clearing and improvement of resolution